# ChraReport # 2012000304
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : **2012000304** PART NUMBER : C1505-B5032 COLUMN TYPE : ChraSil B100 C18, 150 x 4.6 mm, 5.0um LOT : VD184003 | **TEST CONDITIONS** MOBILE PHASE : 85% Acetonitrile / 15% water COLUMN PRESSURE : 35Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (normally 23 0C) INJECTTION VOLUME : 3.0 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 17455 MIN : 16780 SELECTIVITY : 3.27 RANGE : 3.02 - 3.62 USP TAILING FACTOR : 1.13 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.89 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2012000304 "2012000304") at https://chracert.com/2012000304 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : **2012000314** PART NUMBER : C1505-B5034 COLUMN TYPE : ChraSil B100 C18, 250 x 4.6 mm, 5.0um LOT : KL0051 | **TEST CONDITIONS** MOBILE PHASE : 80% Acetonitrile / 20% water COLUMN PRESSURE : 67Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (normally 23 0C) INJECTTION VOLUME : 0.5 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 23067 MIN : 92268 SELECTIVITY : 3.27 RANGE : 3.02 - 3.62 USP TAILING FACTOR : 1.01 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.21 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2012000314 "2012000314") at https://chracert.com/2012000314 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : 2012000324 PART NUMBER : C1505-B5034 COLUMN TYPE : ChraSil B100 C18, 250 x 4.6 mm, 5.0um LOT : KL146923 | **TEST CONDITIONS** MOBILE PHASE : 65% Acetonitrile / 35% water COLUMN PRESSURE : 50Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0C) INJECTTION VOLUME : 0.5 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 13800 MIN : 91998 SELECTIVITY : 3.27 RANGE : 3.02 - 3.62 USP TAILING FACTOR : 1.05 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 3.39 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2012000324 "2012000324") at https://chracert.com/2012000324 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : 2002000214 PART NUMBER : C1515-B5032 COLUMN TYPE : ChraSil B100 C8, 150 x 4.6 mm, 5.0um LOT : KL0243 | **TEST CONDITIONS** MOBILE PHASE : 65% Acetonitrile / 35% water COLUMN PRESSURE : 50Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0C) INJECTTION VOLUME : 0.5 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 13300 MIN : 91398 SELECTIVITY : 3.27 RANGE : 3.02 - 3.62 USP TAILING FACTOR : 1.05 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 3.22 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2002000214 "2002000214") at https://chracert.com/2002000214 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : 2002000204 PART NUMBER : C1515-B5032 COLUMN TYPE : ChraSil B100 C8, 150 x 4.6 mm, 5.0um LOT : KL0051 | **TEST CONDITIONS** MOBILE PHASE : 65% Acetonitrile / 35% water COLUMN PRESSURE : 50Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0C) INJECTTION VOLUME : 0.5 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 13800 MIN : 91998 SELECTIVITY : 3.27 RANGE : 3.02 - 3.62 USP TAILING FACTOR : 1.05 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 3.39 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2002000204 "2002000204") at https://chracert.com/2002000204 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : 2002000174 PART NUMBER : C1505-B5034 COLUMN TYPE : ChraSil B100 C18, 250 x 4.6 mm, 5.0um LOT : KL0051 | **TEST CONDITIONS** MOBILE PHASE : 80% Acetonitrile / 20% water COLUMN PRESSURE : 67Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0C) INJECTTION VOLUME : 0.5 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 22948 MIN : 91793 SELECTIVITY : 3.27 RANGE : 3.02 - 3.62 USP TAILING FACTOR : 1.01 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.21 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2002000174 "2002000174") at https://chracert.com/2002000174 |
CERTIFICATE OF ANALYSIS **QUALITY CONTROL PERFORMANCE REPORT** | |||
**PRODUCT INFORMATION** SERIAL NUMBER : 2019006133 PART NUMBER : S8034-034360 PRODUCT TYPE : ChraSep SPE Columns ODS-IC 500mg/3ml, 50/pk LOT : 639753201 | [](https://chratec.info/uploads/images/gallery/2021-08/spe1.png) | ||
RAW MATERIALS | |||
Component Properties | |||
Properties | Specifications | Results | Methods |
Tube Purity | Proprietary | Pass | GC FID Test |
Frit Purity | Proprietary | Pass | GC FID Test |
Base Silica Physical Properties | |||
Properties | Specifications | Results | Methods |
Surface Area (m2/g) | 450-550 | 535 | N2 BET |
Particle Size (µm) | 45-60 | 46 | Malvern 100 |
Pore Diameter (Å) | 60-90 | N2 BET | |
Particle Size Distribution (%) | ≤20µm ≤ 35µm ≥80µm | ≤20µm ≤ 35µm ≥80µm | Malvern 100 |
≤ 11 ≤18 ≤15 | ≤ 8 ≤15 ≤10 | ||
PRODUCT SPECIFICATION/ANALYSIS | |||
Bonded Silica Properties | |||
Properties | Specifications | Results | Methods |
Carbon Loading (%C) | 17.0-20.0 | 19.5 | CHNO-S Analysis |
Hydrogen Loading(%H) | 1.5-6.0 | 3.95 | CHNO-S Analysis |
Surface Coverage (µmol/m2) | 1.90-2.50 | 2.05 | Calculated based on %C |
Turbidity (NTU) | ≤10.0 | 0.55 | Turbidity meter |
Washable Residue (mg/g) | ≤2.0 | 0.2 | Methanol and Hexane gravimetric |
Bed Mass Consistency | Proprietary | Pass | Weight Measurement |
Flow Characteristics | Proprietary | Pass | Methanol in low vacuum (1” Hg) |
HPLC Test Conditions | |||
Column Dimension | 100 x 4.6 mm | ||
Flow | 1 mL/min | ||
Detector | UV@254 nm | ||
Mobile Phase | Methanol:Water (60:40) | ||
Temperature | 25-35 oC | ||
Probes | Uracil, Acetophenone, Nitrobenzene, Benzene, Toluene | ||
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |||
[ChraReport](https://chrareport.com/2019006133 "2019006133") at https://chracert.com/2019006133 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : 2002000164 PART NUMBER : C1505-B5034 COLUMN TYPE : ChraSil B100 C18, 250 x 4.6 mm, 5.0um LOT : KL0051 | **TEST CONDITIONS** MOBILE PHASE : 80% Acetonitrile / 20% water COLUMN PRESSURE : 67Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0C) INJECTTION VOLUME : 0.5 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 23067 MIN : 92268 SELECTIVITY : 3.27 RANGE : 3.02 - 3.62 USP TAILING FACTOR : 1.01 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.21 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2002000164 "2002000164") at https://chracert.com/2002000164 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : 2002000144 PART NUMBER : C1505-B5034 COLUMN TYPE : ChraSil B100 C18, 250 x 4.6 mm, 5.0um LOT : KL0051 | **TEST CONDITIONS** MOBILE PHASE : 80% Acetonitrile / 20% water COLUMN PRESSURE : 67Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0C) INJECTTION VOLUME : 0.5 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 23452 MIN : 93808 SELECTIVITY : 3.27 RANGE : 3.02 - 3.62 USP TAILING FACTOR : 1.05 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.22 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2002000144 "2002000144") at https://chracert.com/2002000144 |
#### **Certificate of Quality** | ||||||
Item | Material | Description | Qty | QC Test | Origin | |
1 | \[P4819-01479\] | ChraSep, Nonsterile Hydrophilic PTFE Syringe Filters, 0.45(µm), 13(mm), 100pcs | 100 | PASSED | CN | |
2 | ||||||
3 | ||||||
WE HEREBY GUARANTEE THAT THE COMMODITY HEREOF IS MADE OF THE BEST MATERIALS WITH FIRST CLASS WORKMANSHIP, BRAND NEW AND UNUSED AND COMPLIES IN ALL RESPECTS WITH THE QUALITY AS STIPULATED IN THE CONTRACT AND UNDER ISO-9001 QUALITY CONTROL AND VALIDATION OF OBSERVATIONS | ||||||
Signed by: ChraWork | ||||||
Representative Date: 29-May-2020 | ||||||
ChraReport at https://chracert.com/2110000333 | ||||||
#### **Certificate of Quality** | ||||||
Item | Material | Description | Qty | QC Test | Origin | |
1 | \[P4819-01469\] | ChraSep, Nonsterile Hydrophilic PTFE Syringe Filters, 0.22(µm), 13(mm), 100pcs | 100 | PASSED | CN | |
2 | ||||||
3 | ||||||
WE HEREBY GUARANTEE THAT THE COMMODITY HEREOF IS MADE OF THE BEST MATERIALS WITH FIRST CLASS WORKMANSHIP, BRAND NEW AND UNUSED AND COMPLIES IN ALL RESPECTS WITH THE QUALITY AS STIPULATED IN THE CONTRACT AND UNDER ISO-9001 QUALITY CONTROL AND VALIDATION OF OBSERVATIONS | ||||||
Signed by: ChraWork | ||||||
Representative Date: 2-Nov-2020 | ||||||
ChraReport at https://chracert.com/2110000343 | ||||||
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : 2110110343 PART NUMBER : E07219-C4126 COLUMN TYPE : ChraSil Inert HILIC, 150 x 2.1 mm, 3.5um LOT : B08004 | **TEST CONDITIONS** MOBILE PHASE : 75% Acetonitrile / 25% water COLUMN PRESSURE : 169.6Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.172 cm / sec TEMPERATURE : AMBIENT (normally 23 0C) INJECTTION VOLUME : 5.0 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 23184 MIN : 18000 SELECTIVITY : 1.66 RANGE : 1.62 - 1.72 USP TAILING FACTOR : 1.04 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.05 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2110110343 "2110110343") at https://chracert.com/2110110343 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : 2110120343 PART NUMBER : E07219-C4126 COLUMN TYPE : ChraSil Inert HILIC, 150 x 2.1 mm, 3.5um LOT : B08004 | **TEST CONDITIONS** MOBILE PHASE : 75% Acetonitrile / 25% water COLUMN PRESSURE : 169 Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.172 cm / sec TEMPERATURE : AMBIENT (normally 23 0C) INJECTTION VOLUME : 5.0 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 23281 MIN : 18200 SELECTIVITY : 1.63 RANGE : 1.60 - 1.72 USP TAILING FACTOR : 1.06 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.09 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2110110343 "2110110343") at https://chracert.com/2110120343 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : 2110220343 PART NUMBER : E07219-C4126 COLUMN TYPE : ChraSil Inert HILIC, 150 x 2.1 mm, 3.5um LOT : B08004 | **TEST CONDITIONS** MOBILE PHASE : 75% Acetonitrile / 25% water COLUMN PRESSURE : 169 Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.172 cm / sec TEMPERATURE : AMBIENT (normally 23 0C) INJECTTION VOLUME : 5.0 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 22281 MIN : 18270 SELECTIVITY : 1.69 RANGE : 1.60 - 1.72 USP TAILING FACTOR : 1.04 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.00 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2110110343 "2110110343") at https://chracert.com/2110220343 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : 2110420343 PART NUMBER : C1513-B3003 COLUMN TYPE : ChraSil B100 C8, 150 x 2.1 mm, 3.5um LOT : KL0051 | **TEST CONDITIONS** MOBILE PHASE : 75% Acetonitrile / 25% water COLUMN PRESSURE : 145 Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.172 cm / sec TEMPERATURE : AMBIENT (normally 23 0C) INJECTTION VOLUME : 1.0 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 20080 MIN : 18110 SELECTIVITY : 1.69 RANGE : 1.60 - 1.72 USP TAILING FACTOR : 1.04 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 1.60 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2110110343 "2110110343") at https://chracert.com/2110420343 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : 2110520343 PART NUMBER : C1513-B3003 COLUMN TYPE : ChraSil B100 C8, 150 x 2.1 mm, 3.5um LOT : KL0051 | **TEST CONDITIONS** MOBILE PHASE : 75% Acetonitrile / 25% water COLUMN PRESSURE : 145 Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.172 cm / sec TEMPERATURE : AMBIENT (normally 23 0C) INJECTTION VOLUME : 1.0 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 21680 MIN : 18310 SELECTIVITY : 1.60 RANGE : 1.60 - 1.72 USP TAILING FACTOR : 1.03 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 1.72 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2110110343 "2110110343") at https://chracert.com/2110520343 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : 2207000362 PART NUMBER : C1505-B5032 COLUMN TYPE : ChraSil B100 C18, 150 x 4.6 mm, 5.0um LOT : KL0051 | **TEST CONDITIONS** MOBILE PHASE : 80% Acetonitrile / 20% water COLUMN PRESSURE : 4.4 MPa COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0C) INJECTTION VOLUME : 0.5 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 109277 MIN : 16392 SELECTIVITY : 2.17 RANGE : 3.02 - 3.62 USP TAILING FACTOR : 1.05 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.22 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraCert](https://chracert.com/2207000362 "2207000362") at https://chracert.com/2207000362 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : **2307000373** PART NUMBER : C1456-A3507 COLUMN TYPE : ChraSil D300 C18 3.5 um, 100 x 3.0 mm LOT : B13223 | **TEST CONDITIONS** MOBILE PHASE : 60% Acetonitrile / 40% water COLUMN PRESSURE : 175Bar COLUMN FLOW : 0.50 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (normally 23 0C) INJECTTION VOLUME : 3.0 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 17455 MIN : 16780 SELECTIVITY : 3.27 RANGE : 3.02 - 3.62 USP TAILING FACTOR : 1.13 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.89 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
ChraReport at https://chracert.com/2307000373 |
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : **2310000381** PART NUMBER : C0902-A3509 COLUMN TYPE : ChraSil A100 C18 3.5 um, 150 x 3.0 mm LOT : B13223 | **TEST CONDITIONS** MOBILE PHASE : 70% Acetonitrile / 30% water COLUMN PRESSURE : 324Bar COLUMN FLOW : 0.80 ml / min LINEAR VELOCITY : 0.343 cm / sec TEMPERATURE : AMBIENT (normally 23 0C) INJECTTION VOLUME : 2.0 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 32858 MIN : 30000 SELECTIVITY : 1.82 RANGE : 1.78 - 1.88 USP TAILING FACTOR : 0.99 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.26 | Peak Conc Sample (µg/ml) Component 1. 10 Uracil 2. 400 Phenol 3. 50 4-Chloro Nitrobenzene 4. 80 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
ChraReport at https://chracert.com/2310000381 |