ChraReport Certified document and report made by Chratec 2012000304 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2012000304 PART NUMBER : C1505-B5032 COLUMN TYPE : ChraSil B100 C18, 150 x 4.6 mm, 5.0um LOT : VD184003 TEST CONDITIONS MOBILE PHASE : 85% Acetonitrile / 15% water COLUMN PRESSURE : 35Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (normally 23 0 C) INJECTTION VOLUME  : 3.0 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 17455       MIN  : 16780 SELECTIVITY                : 3.27          RANGE : 3.02 - 3.62 USP TAILING FACTOR  : 1.13          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 2.89 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2012000304 2012000314 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2012000314 PART NUMBER : C1505-B5034 COLUMN TYPE : ChraSil B100 C18, 250 x 4.6 mm, 5.0um LOT : KL0051 TEST CONDITIONS MOBILE PHASE : 80% Acetonitrile / 20% water COLUMN PRESSURE : 67Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (normally 23 0 C) INJECTTION VOLUME  : 0.5 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 23067       MIN  : 92268 SELECTIVITY                : 3.27          RANGE : 3.02 - 3.62 USP TAILING FACTOR  : 1.01          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 2.21 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2012000314 2012000324 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2012000324 PART NUMBER : C1505-B5034 COLUMN TYPE : ChraSil B100 C18, 250 x 4.6 mm, 5.0um LOT : KL146923 TEST CONDITIONS MOBILE PHASE : 65% Acetonitrile / 35% water COLUMN PRESSURE : 50Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0 C) INJECTTION VOLUME  : 0.5 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 13800       MIN  : 91998 SELECTIVITY                : 3.27          RANGE : 3.02 - 3.62 USP TAILING FACTOR  : 1.05          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 3.39 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2012000324 2002000214 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2002000214 PART NUMBER : C1515-B5032 COLUMN TYPE : ChraSil B100 C8, 150 x 4.6 mm, 5.0um LOT : KL0243 TEST CONDITIONS MOBILE PHASE : 65% Acetonitrile / 35% water COLUMN PRESSURE : 50Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0 C) INJECTTION VOLUME  : 0.5 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 13300       MIN  : 91398 SELECTIVITY                : 3.27          RANGE : 3.02 - 3.62 USP TAILING FACTOR  : 1.05          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 3.22 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2002000214 2002000204 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2002000204 PART NUMBER : C1515-B5032 COLUMN TYPE : ChraSil B100 C8, 150 x 4.6 mm, 5.0um LOT : KL0051 TEST CONDITIONS MOBILE PHASE : 65% Acetonitrile / 35% water COLUMN PRESSURE : 50Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0 C) INJECTTION VOLUME  : 0.5 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 13800       MIN  : 91998 SELECTIVITY                : 3.27          RANGE : 3.02 - 3.62 USP TAILING FACTOR  : 1.05          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 3.39 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2002000204 2002000174 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2002000174 PART NUMBER : C1505-B5034 COLUMN TYPE : ChraSil B100 C18, 250 x 4.6 mm, 5.0um LOT : KL0051 TEST CONDITIONS MOBILE PHASE : 80% Acetonitrile / 20% water COLUMN PRESSURE : 67Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0 C) INJECTTION VOLUME  : 0.5 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 22948       MIN  : 91793 SELECTIVITY                : 3.27          RANGE : 3.02 - 3.62 USP TAILING FACTOR  : 1.01          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 2.21 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2002000174 2019006133 CERTIFICATE OF ANALYSIS QUALITY CONTROL PERFORMANCE REPORT PRODUCT INFORMATION SERIAL NUMBER : 2019006133 PART NUMBER : S8034-034360 PRODUCT TYPE : ChraSep SPE Columns ODS-IC 500mg/3ml, 50/pk LOT : 639753201   RAW MATERIALS Component Properties Properties Specifications Results Methods Tube Purity Proprietary Pass GC FID Test Frit Purity Proprietary Pass GC FID Test Base Silica Physical Properties Properties Specifications Results Methods Surface Area (m 2 /g) 450-550 535 N2 BET Particle Size (µm) 45-60 46 Malvern 100 Pore Diameter (Å) 60-90   N2 BET Particle Size Distribution (%) ≤20µm ≤ 35µm ≥80µm ≤20µm ≤ 35µm ≥80µm Malvern 100   ≤ 11 ≤18 ≤15 ≤ 8 ≤15 ≤10   PRODUCT SPECIFICATION/ANALYSIS Bonded Silica Properties Properties Specifications Results Methods Carbon Loading (%C) 17.0-20.0 19.5 CHNO-S Analysis Hydrogen Loading(%H) 1.5-6.0 3.95 CHNO-S Analysis Surface Coverage (µmol/m 2 ) 1.90-2.50 2.05 Calculated based on %C Turbidity (NTU) ≤10.0 0.55 Turbidity meter Washable Residue (mg/g) ≤2.0 0.2 Methanol and Hexane gravimetric Bed Mass Consistency Proprietary Pass Weight Measurement Flow Characteristics Proprietary Pass Methanol in low vacuum (1” Hg) HPLC Test Conditions Column Dimension 100 x 4.6 mm Flow 1 mL/min Detector UV@254 nm Mobile Phase Methanol:Water (60:40) Temperature 25-35 o C Probes Uracil, Acetophenone, Nitrobenzene, Benzene, Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2019006133 2002000164 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2002000164 PART NUMBER : C1505-B5034 COLUMN TYPE : ChraSil B100 C18, 250 x 4.6 mm, 5.0um LOT : KL0051 TEST CONDITIONS MOBILE PHASE : 80% Acetonitrile / 20% water COLUMN PRESSURE : 67Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0 C) INJECTTION VOLUME  : 0.5 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 23067       MIN  : 92268 SELECTIVITY                : 3.27          RANGE : 3.02 - 3.62 USP TAILING FACTOR  : 1.01          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 2.21 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2002000164 2002000144 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2002000144 PART NUMBER : C1505-B5034 COLUMN TYPE : ChraSil B100 C18, 250 x 4.6 mm, 5.0um LOT : KL0051 TEST CONDITIONS MOBILE PHASE : 80% Acetonitrile / 20% water COLUMN PRESSURE : 67Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0 C) INJECTTION VOLUME  : 0.5 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 23452       MIN  : 93808 SELECTIVITY                : 3.27          RANGE : 3.02 - 3.62 USP TAILING FACTOR  : 1.05          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 2.22 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2002000144 2110000333 Certificate of Quality   Item Material Description Qty QC Test Origin 1 [P4819-01479] ChraSep, Nonsterile Hydrophilic PTFE Syringe Filters, 0.45(µm), 13(mm), 100pcs 100 PASSED CN 2           3           WE HEREBY GUARANTEE THAT THE COMMODITY HEREOF IS MADE OF THE BEST MATERIALS WITH FIRST CLASS WORKMANSHIP, BRAND NEW AND UNUSED AND COMPLIES IN ALL RESPECTS WITH THE QUALITY AS STIPULATED IN THE CONTRACT AND UNDER ISO-9001 QUALITY CONTROL AND VALIDATION OF OBSERVATIONS Signed by: ChraWork   Representative Date: 29-May-2020         ChraReport at https://chracert.com/2110000333                     2110000343 Certificate of Quality   Item Material Description Qty QC Test Origin 1 [P4819-01469] ChraSep, Nonsterile Hydrophilic PTFE Syringe Filters, 0.22(µm), 13(mm), 100pcs 100 PASSED CN 2           3           WE HEREBY GUARANTEE THAT THE COMMODITY HEREOF IS MADE OF THE BEST MATERIALS WITH FIRST CLASS WORKMANSHIP, BRAND NEW AND UNUSED AND COMPLIES IN ALL RESPECTS WITH THE QUALITY AS STIPULATED IN THE CONTRACT AND UNDER ISO-9001 QUALITY CONTROL AND VALIDATION OF OBSERVATIONS Signed by: ChraWork   Representative Date: 2-Nov-2020         ChraReport at https://chracert.com/2110000343                     2110110343 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2110110343 PART NUMBER : E07219-C4126 COLUMN TYPE : ChraSil Inert HILIC, 150 x 2.1 mm, 3.5um LOT : B08004 TEST CONDITIONS MOBILE PHASE : 75% Acetonitrile / 25% water COLUMN PRESSURE : 169.6Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.172 cm / sec TEMPERATURE : AMBIENT (normally 23 0 C) INJECTTION VOLUME  : 5.0 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 23184       MIN  : 18000 SELECTIVITY                : 1.66          RANGE : 1.62 - 1.72 USP TAILING FACTOR  : 1.04          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 2.05 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2110110343 2110120343 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2110120343 PART NUMBER : E07219-C4126 COLUMN TYPE : ChraSil Inert HILIC, 150 x 2.1 mm, 3.5um LOT : B08004 TEST CONDITIONS MOBILE PHASE : 75% Acetonitrile / 25% water COLUMN PRESSURE : 169 Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.172 cm / sec TEMPERATURE : AMBIENT (normally 23 0 C) INJECTTION VOLUME  : 5.0 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 23281       MIN  : 18200 SELECTIVITY                : 1.63          RANGE : 1.60 - 1.72 USP TAILING FACTOR  : 1.06          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 2.09 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2110120343 2110220343 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2110220343 PART NUMBER : E07219-C4126 COLUMN TYPE : ChraSil Inert HILIC, 150 x 2.1 mm, 3.5um LOT : B08004 TEST CONDITIONS MOBILE PHASE : 75% Acetonitrile / 25% water COLUMN PRESSURE : 169 Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.172 cm / sec TEMPERATURE : AMBIENT (normally 23 0 C) INJECTTION VOLUME  : 5.0 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 22281       MIN  : 18270 SELECTIVITY                : 1.69          RANGE : 1.60 - 1.72 USP TAILING FACTOR  : 1.04          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 2.00 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2110220343 2110420343 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2110420343 PART NUMBER : C1513-B3003 COLUMN TYPE : ChraSil B100 C8, 150 x 2.1 mm, 3.5um LOT : KL0051 TEST CONDITIONS MOBILE PHASE : 75% Acetonitrile / 25% water COLUMN PRESSURE : 145 Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.172 cm / sec TEMPERATURE : AMBIENT (normally 23 0 C) INJECTTION VOLUME  : 1.0 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 20080       MIN  : 18110 SELECTIVITY                : 1.69          RANGE : 1.60 - 1.72 USP TAILING FACTOR  : 1.04          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 1.60 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2110420343 2110520343 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2110520343 PART NUMBER : C1513-B3003 COLUMN TYPE : ChraSil B100 C8, 150 x 2.1 mm, 3.5um LOT : KL0051 TEST CONDITIONS MOBILE PHASE : 75% Acetonitrile / 25% water COLUMN PRESSURE : 145 Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.172 cm / sec TEMPERATURE : AMBIENT (normally 23 0 C) INJECTTION VOLUME  : 1.0 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 21680       MIN  : 18310 SELECTIVITY                : 1.60          RANGE : 1.60 - 1.72 USP TAILING FACTOR  : 1.03          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 1.72 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2110520343 2207000362 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2207000362 PART NUMBER : C1505-B5032 COLUMN TYPE : ChraSil B100 C18, 150 x 4.6 mm, 5.0um LOT : KL0051 TEST CONDITIONS MOBILE PHASE : 80% Acetonitrile / 20% water COLUMN PRESSURE : 4.4 MPa COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0 C) INJECTTION VOLUME  : 0.5 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 109277       MIN  : 16392 SELECTIVITY                :  2.17           RANGE : 3.02 - 3.62 USP TAILING FACTOR  : 1.05            RANGE : 0.98 - 1.25 (@ 5% Peak Height)                        k’ : 2.22 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraCert at https://chracert.com/2207000362 2307000373 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2307000373 PART NUMBER : C1456-A3507 COLUMN TYPE : ChraSil D300 C18 3.5 um, 100 x 3.0 mm LOT : B13223 TEST CONDITIONS MOBILE PHASE : 60% Acetonitrile / 40% water COLUMN PRESSURE : 175Bar COLUMN FLOW : 0.50 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (normally 23 0 C) INJECTTION VOLUME  : 3.0 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 17455       MIN  : 16780 SELECTIVITY                : 3.27          RANGE : 3.02 - 3.62 USP TAILING FACTOR  : 1.13          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 2.89 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2307000373 2310000381 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2310000381 PART NUMBER : C0902-A3509 COLUMN TYPE : ChraSil A100 C18 3.5 um, 150 x 3.0 mm LOT : B13223 TEST CONDITIONS MOBILE PHASE : 70% Acetonitrile / 30% water COLUMN PRESSURE : 324Bar COLUMN FLOW : 0.80 ml / min LINEAR VELOCITY : 0.343 cm / sec TEMPERATURE : AMBIENT (normally 23 0 C) INJECTTION VOLUME  : 2.0 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 32858       MIN  : 30000 SELECTIVITY                : 1.82          RANGE : 1.78 - 1.88 USP TAILING FACTOR  : 0.99          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 2.26 Peak                   Conc Sample (µg/ml)           Component 1.                                 10                                         Uracil 2.                                 400                                   Phenol 3.                                 50             4-Chloro Nitrobenzene 4.                                 80                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2310000381