Chratec

Product Specification Sheet by ChraWork a Gimium™ Ecosystem

2002000214

LC COLUMN
QUALITY CONTROL PERFORMANCE REPORT

COLUMN INFORMATION

SERIAL NUMBER : 2002000214

PART NUMBER : C1515-B5032
COLUMN TYPE : ChraSil B100 C8, 150 x 4.6 mm, 5.0um
LOT : KL0243

TEST CONDITIONS
MOBILE PHASE : 65% Acetonitrile / 35% water
COLUMN PRESSURE : 50Bar
COLUMN FLOW : 1.00 ml / min
LINEAR VELOCITY : 0.205 cm / sec
TEMPERATURE : AMBIENT (~ 23 0C)
INJECTTION VOLUME  : 0.5 µl
TEST VALUES                                        SPECIFICATIONS
THEORETICAL PLATES : 13300       MIN  : 91398
SELECTIVITY                : 3.27          RANGE : 3.02 - 3.62
USP TAILING FACTOR  : 1.05          RANGE : 0.98 - 1.25
(@ 5% Peak Height)                       k’ : 3.22
Peak                   Conc Sample (µg/ml)           Component
1.                                 5                                         Uracil
2.                                 200                                   Phenol
3.                                 25             4-Chloro Nitrobenzene
4.                                 850                                  Toluene

skd.jpg

MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST
ChraReport at https://chracert.com/2002000214

Revision #5
Created 9 August 2021 13:18:14 by Admin
Updated 6 October 2021 04:25:15 by Admin