Chratec

Product Specification Sheet by ChraWork a Gimium™ Ecosystem

2012000314

LC COLUMN
QUALITY CONTROL PERFORMANCE REPORT

COLUMN INFORMATION

SERIAL NUMBER : 2012000314

PART NUMBER : C1505-B5034
COLUMN TYPE : ChraSil B100 C18, 250 x 4.6 mm, 5.0um
LOT : KL0051

TEST CONDITIONS
MOBILE PHASE : 80% Acetonitrile / 20% water
COLUMN PRESSURE : 67Bar
COLUMN FLOW : 1.00 ml / min
LINEAR VELOCITY : 0.205 cm / sec
TEMPERATURE : AMBIENT (normally 23 0C)
INJECTTION VOLUME  : 0.5 µl
TEST VALUES                                        SPECIFICATIONS
THEORETICAL PLATES : 23067       MIN  : 92268
SELECTIVITY                : 3.27          RANGE : 3.02 - 3.62
USP TAILING FACTOR  : 1.01          RANGE : 0.98 - 1.25
(@ 5% Peak Height)                       k’ : 2.21
Peak                   Conc Sample (µg/ml)           Component
1.                                 5                                         Uracil
2.                                 200                                   Phenol
3.                                 25             4-Chloro Nitrobenzene
4.                                 850                                  Toluene

skd.jpg

MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST
ChraReport at https://chracert.com/2012000314

Revision #7
Created 9 August 2021 09:43:05 by Admin
Updated 6 October 2021 04:24:33 by Admin