2012000314 LC COLUMN QUALITY CONTROL PERFORMANCE REPORT COLUMN INFORMATION SERIAL NUMBER : 2012000314 PART NUMBER : C1505-B5034 COLUMN TYPE : ChraSil B100 C18, 250 x 4.6 mm, 5.0um LOT : KL0051 TEST CONDITIONS MOBILE PHASE : 80% Acetonitrile / 20% water COLUMN PRESSURE : 67Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (normally 23 0 C) INJECTTION VOLUME  : 0.5 µl TEST VALUES                                        SPECIFICATIONS THEORETICAL PLATES : 23067       MIN  : 92268 SELECTIVITY                : 3.27          RANGE : 3.02 - 3.62 USP TAILING FACTOR  : 1.01          RANGE : 0.98 - 1.25 (@ 5% Peak Height)                       k’ : 2.21 Peak                   Conc Sample (µg/ml)           Component 1.                                 5                                         Uracil 2.                                 200                                   Phenol 3.                                 25             4-Chloro Nitrobenzene 4.                                 850                                  Toluene MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST ChraReport at https://chracert.com/2012000314