2110110343
LC COLUMN QUALITY CONTROL PERFORMANCE REPORT |
|
COLUMN INFORMATION SERIAL NUMBER : 2110110343 PART NUMBER : E07219-C4126 |
TEST CONDITIONS MOBILE PHASE : 75% Acetonitrile / 25% water COLUMN PRESSURE : 169.6Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.172 cm / sec TEMPERATURE : AMBIENT (normally 23 0C) INJECTTION VOLUME : 5.0 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 23184 MIN : 18000 SELECTIVITY : 1.66 RANGE : 1.62 - 1.72 USP TAILING FACTOR : 1.04 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.05 |
Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
ChraReport at https://chracert.com/2110110343 |