2207000362
LC COLUMN QUALITY CONTROL PERFORMANCE REPORT |
|
COLUMN INFORMATION SERIAL NUMBER : 2207000362 PART NUMBER : C1505-B5032 |
TEST CONDITIONS MOBILE PHASE : 80% Acetonitrile / 20% water COLUMN PRESSURE : 4.4 MPa COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0C) INJECTTION VOLUME : 0.5 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 109277 MIN : 16392 SELECTIVITY : 2.17 RANGE : 3.02 - 3.62 USP TAILING FACTOR : 1.05 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.22 |
Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
ChraCert at https://chracert.com/2207000362 |