2207000362

LC COLUMN
QUALITY CONTROL PERFORMANCE REPORT

COLUMN INFORMATION

SERIAL NUMBER : 2207000362

PART NUMBER : C1505-B5032
COLUMN TYPE : ChraSil B100 C18, 150 x 4.6 mm, 5.0um
LOT : KL0051

TEST CONDITIONS
MOBILE PHASE : 80% Acetonitrile / 20% water
COLUMN PRESSURE : 4.4 MPa
COLUMN FLOW : 1.00 ml / min
LINEAR VELOCITY : 0.205 cm / sec
TEMPERATURE : AMBIENT (~ 23 0C)
INJECTTION VOLUME  : 0.5 µl
TEST VALUES                                        SPECIFICATIONS
THEORETICAL PLATES : 109277       MIN  : 16392
SELECTIVITY                :  2.17           RANGE : 3.02 - 3.62
USP TAILING FACTOR  : 1.05            RANGE : 0.98 - 1.25
(@ 5% Peak Height)                        k’ : 2.22
Peak                   Conc Sample (µg/ml)           Component
1.                                 5                                         Uracil
2.                                 200                                   Phenol
3.                                 25             4-Chloro Nitrobenzene
4.                                 850                                  Toluene