**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT**
**COLUMN INFORMATION** SERIAL NUMBER : 2207000362 PART NUMBER : C1505-B5032 COLUMN TYPE : ChraSil B100 C18, 150 x 4.6 mm, 5.0um LOT : KL0051 **TEST CONDITIONS** MOBILE PHASE : 80% Acetonitrile / 20% water COLUMN PRESSURE : 4.4 MPa COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (~ 23 0C) INJECTTION VOLUME : 0.5 µl
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 109277 MIN : 16392 SELECTIVITY : 2.17 RANGE : 3.02 - 3.62 USP TAILING FACTOR : 1.05 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.22Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene